mass_G_100

technical parameter
liquid
Measurement Principle: Coriolis Force Mass Flow Measurement
Product features: compact high-pressure sensor, with ultra-compact integrated transmitter. High-precision liquid and gas measurements are carried out in high-pressure applications. Sensor features simple and safe process integration: threaded connection
Less number of process measurement points: multi-variables (flow, density, temperature)
The required installation space is small: the length of straight pipe section without front and rear requires the maximum process pressure of connection during threading process to be 350 bar (5080 psi)
Transmitter features: small installation controls required: in the smallest space to achieve all functions save time on-site operation, without other software and hardware: built-in Web server
Built-in Validation Function: Heartbeat Technology Hard and Durable Ultra-compact Transmitter Shell Pre-set Connection Plug
Nominal Caliber Range: DN 8... 25 (niche... 1")
Material for liquid receiving parts: measuring tube: 1.4435 (316L)
Process Connection: 1.4404 (316/316L)
Measurement variables: mass flow, density, temperature, volume flow, corrected volume flow, reference density
 
Maximum measurement error: mass flow (liquid): +0.15%
Volume flow (liquid): +0.15%
Mass Flow (Gas): +0.75%
Density (liquid): +0.0005 g/cm_
Measurement range: 0.18 000 kg/h (0.662 lb/min)
Maximum process pressure: 350 bar (5080 psi)
Medium temperature range: - 50... + 150 degree C (- 58... + 302 degree F)
Ambient temperature range: - 40... + 60 degree C (- 40... + 140 degree F)
Optional: - 50... + 60 degrees C (- 58... + 140 degrees F)
Sensor shell material: 1.4301 (304), corrosion resistance
Transmitter Shell Material in One Form: AlSi10Mg Aluminum Alloy with Coating
One-in-one/ultra-compact one-in-one: 1.4301 (304); 1.4404 (316L)
Protection Level: IP66/67, type 4X (Shell)
Display/operation: Four-line backlight display (no field operation) is set through Web server and debugging tools
Output: 4... 20 mA HART (Active) Pulse/Frequency/Switch Output (Passive)
Input: None
Digital communication: HART, Modbus RS485, Industrial Ethernet (EtherNet/IP), PROFIBUS DP
Power supply: 20... 30 V DC
Risk zone certification: ATEX, IECEx, cCSAus, NEPSI, INMETRO
Other Certificates and Certificates: 3.1 Material Certificate, Performing Calibration on Calibration Device (Complying with ISO/IEC 17025 Standard)