Consultation Hotline:0755-83990550
Ch/En
Localhost:Home > E+H > flow > Vortex Flowmeter >

Prowirl_D_200

Gas
Measurement Principle: Vortex Street Flow Measurement
Product features: Flowmeter adopts economical clamping design, providing integrated or split instruments. Built-in temperature measurement is suitable for mass and energy measurement of saturated steam. Suitable for all basic applications, as well as with 1 pair of 1 replacement orifice plate occasions.
Sensor Characteristics: Easy to align sensor: High applicability with middle ring: Strong structure, anti-vibration, temperature shock-water hammer without maintenance: Life calibration installation length of 65 mm (2.56 in) can not be blue.
Transmitter Features: Instrument Wiring Convenience: Independent Cavity Operation Safety: Backlight Display Unit with Touch Keys, No Instrument Opening
Built-in Verification: Heartbeat Display Module with Data Output Function Strong Dual-chamber Shell
Nominal caliber range: DN 15... 150 (... 6") Material measuring tube for liquid receiving parts: 1. 4408 (C3FM) DSC sensor: 1. 4435 (316/316L)
Measuring variables: volume flow, mass flow, corrected volume flow, energy flow, difference of heat flow, maximum error of temperature measurement volume flow (liquid): +0.75% volume flow (steam, gas): +1.00% mass flow (liquid): +0.85% mass flow (steam, gas): +1.7% measurement range liquid: 0.16... 625 M _/h (0.09... 368 ft_/min)
Depending on the medium: water, 1 bar a (14.5 psia), 20 degree C (68 degree F) steam, gas: 2... 8342 m_/h (1.18... 4910 ft_/min)
Depending on the medium: steam, 180 degree C (356 degree F), 10 bar a (145 psia) air, 25 degree C (77 degree F), 4.4 bar a (63.8 psia)
Maximum process pressure: PN 40, Cl. 300, 20K
Medium Temperature Range Standard: - 40... + 260 C (- 40... + 500 F)
HIGH/LOW TEMPERATURE (OPTIONS): - 200... + 400 ~C (- 328... + 752 ~F)
High/Low Temperature (Special Selection): - 200... + 450 C (- 328... + 842 F)
Integrated Instrument for Environmental Temperature Range (Standard): - 40... + 80 C (- 40... + 176 F)
Integrated Instrument (Option): - 50... + 80 C (- 58... + 176 F)
Separate Instrument (Standard): - 40... + 85 C (- 40... + 185 F)
Separate Instruments (Options): - 50... + 85 C (- 58... + 185 F)
Sensor Shell Material Sensor Connection Shell: AlSi10Mg Aluminum Alloy Coating, 1.4408 (CF3M) Transmitter Shell Material AlSi10Mg Aluminum Alloy Coating, 1.4404 (316L)
Integrative Instrument for Protection Level: IP66/67, Type 4X (Shell)
Separate Sensors: IP66/67, Type 4X (Shell)
Separate Transmitter: IP66/67, Type 4X (Shell)
Display/Operation: Four-line backlight display, touch key control (external operation) through the field display unit and debugging tools to set up optional separate display
Output: 4... 20 mA HART (Passive) 4... 20 mA (Passive) Pulse/Frequency/Switch Output (Passive) Input 4... 20 mA (Passive) Digital Communication HART, PROFIBUS PA
Supply Voltage: 12.35 V DC (4.20 mA HART, with/without pulse/frequency/switching output) 12.30 V (4.20 mA HART, 4.20 mA) 12.35 V DC (4.20 mA HART, pulse/frequency/switching output, 4... 20 mA input)
Ex certification: ATEX, IECEx, cCSAus
Other Certificates and Certificates: 3.1 Material Certificate, perform calibration on certified calibration equipment (conforming to ISO/IEC 17025 standard), NAMUR, SIL, PED, CRN, NACE MR0175/MR0103, PMI (special selection)

liquid
Measurement Principle: Vortex Street Flow Measurement
Product features: Flowmeter adopts economical clamping design, providing integrated or split instruments. Built-in temperature measurement is suitable for mass and energy measurement of saturated steam. Suitable for all basic applications, as well as with 1 pair of 1 replacement orifice plate occasions.
Sensor features: easy alignment sensor: with center ring
High Applicability: Verified Structural Firmness, Vibration Resistance, Temperature Impact-Water Hammer
Maintenance Free: Life Calibration Installation Length 65 mm (2.56 in) Can't Blue
Transmitter Characteristics: Instrument Wiring Convenience: Independent Wiring Cavity
Operational Safety: Backlight Display Unit with Touch Keys, No Instrument Opening
Built-in Verification: Heartbeat Display Module with Data Output Function Strong Dual-chamber Shell
Nominal Caliber Range: DN 15... 150 (... 6")
Material of liquid receiving parts: measuring tube: 1.4408 (C3FM) DSC sensor: 1.4435 (316/316L)
Measurement variables: volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
Maximum measurement error: volume flow (liquid): +0.75% volume flow (steam, gas): +1.00% mass flow (liquid): +0.85% mass flow (steam, gas): +1.7%.
Measuring range liquid: 0.16... 625 m_/h (0.09... 368 ft_/min)
Depending on the medium: water, 1 bar a (14.5 psia), 20 degree C (68 degree F) steam, gas: 2... 8342 m_/h (1.18... 4910 ft_/min)
Depending on the medium: steam, 180 degree C (356 degree F), 10 bar a (145 psia) air, 25 degree C (77 degree F), 4.4 bar a (63.8 psia)
Maximum process pressure: PN 40, Cl. 300, 20K
Medium Temperature Range Standard: - 40... + 260 C (- 40... + 500 F)
HIGH/LOW TEMPERATURE (OPTIONS): - 200... + 400 ~C (- 328... + 752 ~F)
High/Low Temperature (Special Selection): - 200... + 450 C (- 328... + 842 F)
Environmental Temperature Range: Integrated Instrument (Standard): - 40... + 80 degree C (- 40... + 176 degree F)
Integrated Instrument (Option): - 50... + 80 C (- 58... + 176 F)
Separate Instrument (Standard): - 40... + 85 C (- 40... + 185 F)
Separate Instruments (Options): - 50... + 85 C (- 58... + 185 F)
Sensor Shell Material Sensor Connection Shell: AlSi10Mg Aluminum Alloy Coating, 1.4408 (CF3M)
Aluminum Alloy AlSi10Mg Coating, 1.4404 (316L) for Transmitter Shell Material
Protection Level: Integrated Instrument: IP66/67, Type 4X (Shell)
Separate Sensors: IP66/67, Type 4X (Shell)
Separate Transmitter: IP66/67, Type 4X (Shell)
Display/Operation: Four-line backlight display, touch key control (external operation) through the field display unit and debugging tools to set up optional separate display
Output: 4... 20 mA HART (Passive) 4... 20 mA (Passive) Pulse/Frequency/Switch Output (Passive)
Input: 4... 20 mA (passive) digital communication HART, PROFIBUS PA
Supply Voltage: 12.35 V DC (4.20 mA HART, with/without pulse/frequency/switching output) 12.30 V (4.20 mA HART, 4.20 mA) 12.35 V DC (4.20 mA HART, pulse/frequency/switching output, 4... 20 mA input) 9.32 V (PROFIBUS PA, Pulse/Frequency/Switch Output)
Ex certification: ATEX, IECEx, cCSAus
Other Certificates and Certificates: 3.1 Material Certificate, perform calibration on certified calibration equipment (conforming to ISO/IEC 17025 standard), NAMUR, SIL, PED, CRN, NACE MR0175/MR0103, PMI (special selection)
 
steam
Measurement Principle: Vortex Street Flow Measurement
Product features: Flowmeter adopts economical clamping design, providing integrated or split instruments. Built-in temperature measurement is suitable for mass and energy measurement of saturated steam. Suitable for all basic applications, as well as with 1 pair of 1 replacement orifice plate occasions.
Sensor features: easy alignment sensor: with center ring
High Applicability: Verified Structural Firmness, Vibration Resistance, Temperature Impact-Water Hammer
Maintenance Free: Life Calibration Installation Length 65 mm (2.56 in) Can't Blue
Nominal Caliber Range: DN 15... 150 (... 6")
Measuring range liquid: 0.16... 625 m_/h (0.09... 368 ft_/min)
Depending on the medium: water, 1 bar a (14.5 psia), 20 degrees C (68 degrees F)
Steam and gas: 2... 8342 m_/h (1.18... 4910 ft_/min)
Depending on the medium: steam, 180 degree C (356 degree F), 10 bar a (145 psia) air, 25 degree C (77 degree F), 4.4 bar a (63.8 psia)
Medium Temperature Range: Standard: - 40... + 260 degree C (- 40... + 500 degree F)
HIGH/LOW TEMPERATURE (OPTIONS): - 200... + 400 ~C (- 328... + 752 ~F)
High/Low Temperature (Special Selection): - 200... + 450 C (- 328... + 842 F)
Maximum process pressures PN 40, Cl. 300, 20K
Maximum measurement error: volume flow (liquid): +0.75%
Volume flow (steam, gas): +1.00%
Mass flow rate (liquid): +0.85%
Mass flow rate (steam, gas): +1.7%
Digital Communication: HART, PROFIBUS PA
Output: 4... 20 mA HART (Passive) 4... 20 mA (Passive) Pulse/Frequency/Switch Output (Passive)
Input: 4... 20 mA (passive)
Ex certification: ATEX, IECEx, cCSAus
Other Certificates and Certificates: 3.1 Material Certificate, perform calibration on certified calibration equipment (conforming to ISO/IEC 17025 standard), NAMUR, SIL, PED, CRN, NACE MR0175/MR0103, PMI (special selection)
Protection Level: Integrated Instrument: IP66/67, Type 4X (Shell)
Separate Sensors: IP66/67, Type 4X (Shell)
Separate Transmitter: IP66/67, Type 4X (Shell) Display/Operate Four-Line Backlight Display, Touch Key Control (External Operations) Set Optional Separate Display by Field Display Unit and Debugging Tool