Consultation Hotline:0755-83990550
Ch/En
Localhost:Home > E+H > Water analysis > Ion analysis >

System_CA80FE

technical parameter
Measurement Principle: Spectrophotometric Colorimetric Iron Ion Analyser, Ferrozine Method
Outward Size: Cabinet: 793 x 530 x 417 mm (31.22 x 20.87 x 16.42 in)
Standard case: 1693 x 530 x 417 mm (66.65 x 20.87 x 16.42 in)
Design: Open, Cabinet and Vertical Cabinet
High Performance Plastics: ASA-PC, Coated Steel Plate Base
Process temperature: 4... 40 degree C (41... 104 degree F)
Ambient temperature: 5... 40 degree C (41... 104 degree F)
Process pressure: At atmospheric pressure, the absolute pressure is less than 0.2 bar.
Sample flow rate: min. 5ml/min (0.17 fl. oz/min), homogeneous water solution with low solid content (turbidity < 50 FNU)
Advantages: Easy upgrade to measurement station, up to four Memosens digital sensors, long service life of reagents, low maintenance cost, remote access through digital communication, dual-channel instrumentation, advanced diagnostic functions to support pre-maintenance
Application: monitoring of residual iron ion concentration after precipitation, flocculation and filtration, monitoring of process water rich in iron, quality control of raw water in food industry and quality control of drinking water
Power supply: 100... 120 VAC / 200... 240 VAC + 10%, 24 VDC + 10%, 50 + 1 Hz or 60 + 1.2 Hz
Output: 2 0/4... 20 mA, Modbus RS485/TCP (optional), Ethernet (optional), alarm relay
Input: 1 or 2 measurement channels, 1... 4 digital Memosens sensor input (optional)
Measuring range: 0.05... 2.5 mg/l (ppm) Fe, 0.1... 5 mg/l (ppm) Fe
Consumables: Reagents and standard solution CY80FE and cleaning solution CY800 are needed. Regular maintenance using Maintenance Kit CAV800.