Localhost:Home > E+H > Water analysis > Ion analysis >

System_CA80NO

technical parameter
Measurement Principle: Photometric Colorimetric Nitrite Analyser, Naphthalene Amine Nitrite Determination Method, in line with DIN EN 26777 Standard
Outward Size: Cabinet: 793 x 530 x 417 mm (31.22 x 20.87 x 16.42 in)
Standard case: 1693 x 530 x 417 mm (66.65 x 20.87 x 16.42 in)
Design: Open, Cabinet and Vertical Cabinet
High Performance Plastics: ASA-PC, Coated Steel Plate Base
Process temperature: 4... 40 degree C (41... 104 degree F)
Ambient temperature: 5... 40 degree C (41... 104 degree F)
Process pressure: At atmospheric pressure, the absolute pressure is less than 0.2 bar.
Sample flow rate: min. 5ml/min (0.17 fl. oz/min), homogeneous water solution with low solid content (turbidity < 50 FNU)
Advantages: Easy upgrade to measurement station, up to four Memosens digital sensors, automatic calibration and cleaning, instruments with cooling units can extend the quality guarantee period of reagents, remote access through digital communication, and dual-channel instruments.
Application: nitrite monitoring in drinking water, raw water quality control in food industry, nitrite monitoring in mineral water production process
Power supply: 100... 120 VAC / 200... 240 VAC + 10%, 24 VDC + 10%, 50 + 1 Hz or 60 + 1.2 Hz
Output: 2 0/4... 20 mA, Modbus RS485/TCP (optional), Ethernet (optional), alarm relay
Input: 1 or 2 measurement channels, 1... 4 digital Memosens sensor input (optional)
Measurement range: 10... 500 ug/l NO2-N, 0.1... 1.0 mg/l NO2-N, 0.2... 3.0 mg/l NO2-N
Consumables: Reagents and standard solution CY80NO and cleaning solution CY800 are needed. Regular maintenance using Maintenance Kit CAV800.