technical parameter
Measurement Principle: Spectrophotometric Colorimetric Total Phosphorus Analyzer, Molybdenum Blue Method, in accordance with DIN EN ISO 6878 Standard
Outward Size: Cabinet: 793 x 530 x 417 mm (31.22 x 20.87 x 16.42 in)
Standard case: 1693 x 530 x 417 mm (66.65 x 20.87 x 16.42 in)
Design: Open, Cabinet and Vertical Cabinet
High Performance Plastics: ASA-PC, Coated Steel Plate Base
Process temperature: 4... 40 degree C (41... 104 degree F)
Ambient temperature: 5... 40 degree C (41... 104 degree F)
Process pressure: At atmospheric pressure, the absolute pressure is less than 0.2 bar.
Sample flow rate: min. 5ml/min (0.17 fl. oz/min)
Sample homogeneity: homogeneous aqueous solution with low solid content (turbidity < 50 FNU)
Advantages: Easy upgrade to measurement station, up to four Memosens digital sensors, remote access through digital communication, high-precision input unit, dual-channel instrumentation, advanced diagnostic functions to support pre-maintenance
Application: Sewage treatment plant outlet monitoring, for document recording, sewage treatment plant inlet monitoring, industrial wastewater monitoring, industrial water treatment monitoring
Power supply: 100... 120 VAC / 200... 240 VAC + 10%, 24 VDC + 10%, 50 + 1 Hz or 60 + 1.2 Hz
Output: 2 0/4... 20 mA, Modbus RS485/TCP (optional), Ethernet (optional), alarm relay
Input: 1 or 2 measurement channels, 1... 4 digital Memosens sensor input (optional)
Measurement range: 0.05... 10 mg/l P, 0.5... 50 mg/l P
Consumables: Reagent and standard solution CY80TP and cleaning solution CY800 are needed. Regular maintenance using Maintenance Kit CAV800.