Consultation Hotline:0755-83990550
Ch/En
Localhost:Home > E+H > Location > Ultrasonic >

Prosonic_FDU91F

technical parameter
Continuum/liquid
Measurement Principle: Ultrasound Level Instrument
Characteristic/Application: Separate instrument with on-site housing or top rail mounting housing for installation in control cabinet, the distance between sensor and transmitter is 300 m
Power/communication: four-wire system (HART, PROFIBUS DP)
Measurement accuracy: +/-2mm + 0.17% of measurement distance
Ambient temperature: - 40 degree C... + 105 degree C (- 40 degree F... + 221 degree F)
Process temperature: - 40 C... + 105 C (- 40 F... + 221 F), CIP: 135 C (275 F), within 30 minutes
Process pressure (absolute pressure)/maximum overpressure limit: 0.7 bar... 4 bar abs (10 psi... 58 psi)
Main liquid-receiving parts: 316L (full-welded type: IP68/NEMA 6P)
Process Connection: G/NPT 1 ", DN80/100, ANSI 3" / 4“
Sanitary process connection: Tri-Clamp ISO2852 clamp, DN80 3 1/2 ", DN100 4
Blind zone distance: 0.3 m (1 ft)
Maximum measuring distance: 10 m (33 ft)
Output: Transmitter: 4... 20 mA HART, PROFIBUS DP
Certificate/certification: ATEX, FM, CSA, IECEx, EN 10204-3.1, 3A
Application limitations: possible bubble / strong disturbance: FDU92, tank inspection: FMU95
Components: Transmitter: FMU90, FMU95

technical parameter
Continuum/Solid
Measurement Principle: Ultrasound Level Instrument
Characteristic/Application: Separate instrument with on-site housing or top rail mounting housing for installation in control cabinet, the distance between sensor and transmitter is 300 m
Power/communication: four-wire system (HART, PROFIBUS DP)
Measurement accuracy: +/-2mm + 0.17% of measurement distance
Ambient temperature: - 40 degree C... + 105 degree C (- 40 degree F... + 221 degree F)
Process temperature: - 40 C... + 105 C (- 40 F... + 221 F), CIP: 135 C (275 F), within 30 minutes
Process pressure (absolute pressure)/maximum overpressure limit: 0.7 bar... 4 bar abs (10 psi... 58 psi)
Main liquid-receiving parts: 316L (full-welded type: IP68/NEMA 6P)
Process Connection: G/NPT 1 ", DN80/100, ANSI 3" / 4“
Sanitary process connection: Tri-Clamp ISO2852 clamp, DN 80 (31/2"), DN100 (4")
Blind zone distance: 0.3 m (1 ft)
Maximum measurement distance: 5 m (16 ft), output: transmitter: 4... 20 mA HART
PROFIBUS DP
Certificate/Certification: ATEX, CSA
Options: the second 4... 20 mA output
Component: Transmitter: FMU90