Levelflex_FMP54

technical parameter
Continuum/liquid
Measuring Principle: Guided Wave Radar Level Instrument
Features / applications: Premium equipment, high pressure / high temperature, rod probe, cable probe, coaxial probe, built-in data storage unit, factory pre calibration, reliable measurement: wave surface + foam surface, medium replacement.
Interface measurement: clear liquid-liquid interface, simultaneous measurement of interface height and total liquid level
Power/communication: two-wire system (HART, PROFIBUS PA), four-wire system (HART)
Measurement accuracy: rod probe: +/-2 m m (0.08 in), cable probe (<= 15 m (49 ft): +/-2 mm (0.08 in), cable probe (> 15 m (49 ft): +/-10 mm (0.39 in), coaxial probe: +/-5 mm (0.2 in)
Ambient temperature: - 40 degree C... + 80 degree C (- 40 degree F... + 176 degree F)
Process Temperature: Warming Type (XT): - 196... + 280 C (- 321... + 536 F), High Temperature Type (HT): - 196... + 450 C (- 321... + 842 F), Process Pressure (adiabatic) / Maximum Overpressure Limited Value Vacuum. 400 bar (5800 psi)
Main liquid-receiving parts: rod probe: 304, 316L, Alloy C alloy, ceramics, cable probe: 304, 316, 316L, Alloy C alloy, ceramics, coaxial probe: 304, 316L, Alloy C alloy, ceramics, process connection, thread: G 1/2, MNPT 1/2, flange: ASME 2 "... 4", DN50... DN100, JIS 10K, 63K, Fisher, Masoneilan
Sensor length: rod probe: 10 m (33 ft), cable probe: 45 m (148 ft), coaxial probe: 6 m (20 ft)
Maximum measuring distance: rod probe: 10 m (33 ft); min. DK > 1.6; cable probe: 25 m 30 m (82 ft 98 ft); min. DK > 1.6; 30, m 45 m (98 ft 148 ft); min DK > 1.9; coaxial probe: 6 m (20 ft); min. DK > 1.4
Output: 4... 20 mA HART, PROFIBUS PA, Foundation Fieldbus (FF)
Certificate/certification: ATEX, FM, CSA, CSA C/US, IECEx, NEPSI, WHG spillover protection, SIL, EN10204-3.1, NACE, steam boiler certification
Options: split probe with 3 m (9 ft) cable
Application limitation: horizontal tank: using coaxial probe
Interface measurement: Coaxial system (coaxial probe, bypass tube, guided wave tube) is recommended.
Interface of emulsion layer: FMP55