technical parameter
Gas
Measurement Principle: Differential Pressure
Main features: digital transducer, with metal measuring diaphragm, modular transducer, long-term stability, high static pressure/anti-overload, self-diagnostic function, effectively improving safety, with additional process isolation function
Process temperature: gradient change along the tapping tube temperature
Pressure range: max. 420 bar (max. 6. 091 psi)
Measurement error: 0.075%, platinum type: 0.05%, HART, PROFIBUS PA, Foundation Fieldbus (FF)
Output: 4... 20 mA HART, PROFIBUS PA, Foundation Fieldbus (FF)
Explosion-proof Certification (Ex): ATEX, FM, CSA, NEPSI, IECEx, SIL, NACE, GL Classification Certification, WHG/Overflow Protection
EN10204-3.1
Display/Action: Options
technical parameter
liquid
Measurement Principle: Differential Pressure
Main features: digital transducer, with metal measuring diaphragm, modular transducer, long-term stability, high static pressure/anti-overload, self-diagnostic function, effectively improving safety, with additional process isolation function
Process temperature: gradient change along the tapping tube temperature
Pressure range: max. 420 bar (max. 2. 175 psi)
Measurement error: 0.075%, platinum type: 0.05%, HART, PROFIBUS PA, Foundation Fieldbus (FF)
Output: 4... 20 mA HART, PROFIBUS PA, Foundation Fieldbus (FF)
Explosion-proof Certification (Ex): ATEX, FM, CSA, NEPSI, IECEx, SIL, NACE, GL Classification Certification, WHG/Overflow Protection, EN10204-3.1
Display/Action: Options
technical parameter
pressure
Measurement Principle: Differential Pressure
Features: Digital transmitter, with metal measuring diaphragm, modular transmitter, long-term stability, high static pressure/anti-overload, with additional process isolation function
Power/Communication: 4.20 mA HART: 10.5... 45 V DC; Ex ia: 10.5... 30 V DC, PROFIBUS PA / Foundation Fieldbus (FF): 9... 32 V DC
Measurement accuracy: 0.075%, platinum type: 0.05%.
Long-term stability: 0.05% for range upper limit (URL) and 0.125% for range upper limit (URL) for 5 years
Process temperature: - 40 degree C... 120 C (- 40 F.) 248 degrees F)
Ambient temperature: - 40 C... 85 C (- 40 F... 185 F)
Range of measurement: 10 mbar... 40 bar (0.4 in H_2O... 600 psi)
Maximum Overvoltage Limit: Unilateral: 420 bar (6300 psi)
Process Connection: 1/4-18 NPT, RC1/4“
Output: 4... 20 mA HART, PROFIBUS PA, Foundation Fieldbus (FF)
Certificate/Certification: ATEX, FM, CSA, NEPSI, IECEx, WHG/Overflow Protection, SIL, NACE, GL Classification Certification, EN10204-3.1
Options: HistoROM/M-Dat, four-line digital display, stainless steel or aluminium case, separate case
Characteristics: Diagnostic function
technical parameter
steam
Measurement Principle: Differential Pressure
Main features: digital transducer, with metal measuring diaphragm, modular transducer, long-term stability, high static pressure/anti-overload, self-diagnostic function, effectively improving safety, with additional process isolation function
Process temperature: gradient change along the tapping tube temperature
Pressure range: max. 420 bar (max. 6. 091 psi)
Measurement error: 0.075%, platinum type: 0.05%.
Communication: HART, PROFIBUS PA, Foundation Fieldbus (FF)
Output: 4... 20 mA HART, PROFIBUS PA, Foundation Fieldbus (FF)
Explosion-proof Certification (Ex): ATEX, FM, CSA, NEPSI, IECEx, SIL, NACE, GL Classification Certification, WHG/Overflow Protection, EN10204-3.1
Display/Action: Options