Consultation Hotline:0755-83990550
Ch/En
Localhost:Home > E+H > pressure > Process pressure >

Cerabar_M_PMP51

technical parameter
Continuum/liquid
Measurement Principle: Static Pressure
Features/Applications: Intelligent and reliable pressure transmitter, piezoresistive sensor, process isolation diaphragm with metal welding
Power/Communication: 4.20 mA HART: 11.5... 45 V DC; Ex ia: 11.5... 30 V DC, PROFIBUS PA, Foundation Fieldbus (FF)
Measurement accuracy: 0.15%, platinum type: 0.075%, long-term stability, upper limit of range (URL) 0.1%/year
Ambient temperature: - 40 C... 85 C (- 40 F... 185 F)
Process temperature: - 40 C 125 C (- 40 F 275 F), process pressure (absolute pressure)/
Maximum overvoltage limit: 600 bar (9000 psi)
Process pressure range: 1 bar... 400 bar (15 psi... 6000 psi)
Main liquid-receiving parts: 316L
Process Connection: Thread, Flange (DIN, ANSI, JIS)
Sanitary connection: Tri-Clamp ISO2852 clamp
Maximum measuring distance: 4000m (13.123 ft) H_2O
Output: 4... 20 mA HART, PROFIBUS PA, Foundation Fieldbus (FF)
Certificate/certification: ATEX, CSA, FM, IECEx, TIIS, NACE certificate, SIL, EHEDG, 3A, FDA, testing certificate
Options: On-site display
Application limitation: Measuring unit: When metal is welded and under pressure, differential pressure can be measured by two pressure transmitters (electronic differential pressure measurement). Pay attention to the pressure ratio: static pressure.
technical parameter
pressure
Measurement Principle: Static Pressure
Features: Intelligent and reliable pressure transmitter, piezoresistive sensor, process isolation diaphragm with metal welding
Power/Communication: 4.20 mA HART: 11.5... 45 V DC; Ex ia: 11.5... 30 V DC, PROFIBUS PA / Foundation Fieldbus (FF), 9... 32V DC (Non Ex)
Measurement accuracy: 0.15%
Platinum type: 0.075%, long-term stability, 0.1% of maximum range (URL) per year
Process temperature: - 40 degree C... 130 C (- 40 F.) 266 degrees F, within 1 hour: 150 degrees C (302 degrees F)
Ambient temperature: - 40 C... 85 C (- 40 F... 185 F)
Measuring range: 400m bar.... 400 bar. (15 psi.) 6000 psi)
Maximum overvoltage limit: 600 bar (9000 psi)
Process Connection: Thread: G1/2... G2, MNPT1/2... MNPT2
Flange: DN25... DN80, ASME 1/2 "... 4".
Sanitary Connections: Tri-Clamp, DIN 1851, NEUMO, Varivent N, SMS, DR
Output: 4... 20 mA HART, PROFIBUS PA, Foundation Fieldbus (FF)
Certificate/certification: ATEX, CSA, FM, IECEx, TIIS, NACE certificate, SIEHEDG, 3A, FDA, testing certificate
Options: On-site display